| Catalog Number |
367292 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturing location for this product is nipro.This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.D4.Medical device expiration date: unknown.H4.Device manufacture date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using one bd vacutainer® safety-lok¿ blood collection set, the tube will bounce off the needle, causing the tube to fall and delay the sample processing.The patient had to be stuck again but there was no health impact or consequence reported.
|
| |
|
Event or Problem Description
|
|
It was reported that while using one bd vacutainer® safety-lok¿ blood collection set, the tube will bounce off the needle, causing the tube to fall and delay the sample processing.The patient had to be stuck again but there was no health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received samples or photos for evaluation.Additionally, bd was unable to determine the specific lot number associated with this complaint; therefore, a review of the device history record could not be conducted.However, the dhr of 49 lots from the product concerned (#367292) manufactured in 2024 was investigated, and nothing abnormal was found.This complaint is unable to be confirmed.If additional information is made available, this complaint will be reopened to assess the level of investigation needed.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Search Alerts/Recalls
|