| Catalog Number |
18061002 |
| Medical Device Problem Code |
Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation is completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "the instruments, once the patient had already been anesthetized, were found to be defective and unusable.Our agent had to urgently deliver another instrument, which was withdrawn in an emergency from one of our deposit accounts, causing considerable inconvenience to both customers and significantly lengthening the duration of the operation.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since the device was returned and could not be assembled properly.The device inspection revealed the following: the inner part of the nail handle is highly worn out, which led to local deformation from repetitive friction.Both nail holding screws (worn out one and intact one) could not be inserted through the nail handle as intended.They both jammed at the level of the observed inner deformation.The damages observed on the nail holding screw (lot k953367) and the inner nail handle were most likely mutually caused due to friction and forced assembly overtime.As a consequence, the nail holding screw (lot k175477) could not be inserted anymore in the deformed nail handle, despite being intact.The dimensional inspection confirmed the above.There is a local deformation inside the nail handle, reducing the inner diameter.Therefore, both nail holding screws cannot be inserted anymore.Based on investigation, the root cause was attributed to a user related issue.The devices were most likely not inspected and tested properly before the procedure, leading to unnoticed deformation due to wear on the nail holding screw and nail handle.As a reminder, the stryker instructions for cleaning, sterilization, inspection and maintenance help to determine whether an instrument is in good condition or must be replaced due to excessive wear or obvious defects.The instructions for use also state: « before every operation, ensure that all devices to be used during the operation function correctly with each other ».A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "the instruments, once the patient had already been anesthetized, were found to be defective and unusable.Our agent had to urgently deliver another instrument, which was withdrawn in an emergency from one of our deposit accounts, causing considerable inconvenience to both customers and significantly lengthening the duration of the operation.".
|
| |
|