DEPUY SYNTHES PRODUCTS LLC MINIMAL ACCESS DRIVER,ANGLED; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
|
Back to Search Results |
|
| Catalog Number |
MA-D20-G1 |
| Medical Device Problem Codes |
Appropriate Device Problem Term/Code Not Available (3191); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device had unintended/unexpected disengagement.It was further determined that the device failed pretest for cutter brake assessment.Therefore, the reported condition was confirmed.The assignable root cause was determined to be traced to the user, which is user error.
|
| |
|
Event or Problem Description
|
|
It was reported that the angled driver attachment device was broken.During in-house engineering evaluation it was determined that the device had unintended/unexpected disengagement.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2025.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|