• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL GALAXY G3; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL GALAXY G3; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number GLY120407
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/07/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer ref#: (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section d2b ¿ procode: krd/hcg section h3 - the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
As reported by a health care professional, during a coil embolization to the left main coronary artery (lmca), a non j&j microcatheter was placed and a galaxy g3 mini 2mm x 6cm coil (glm920060, 31136265) was delivered to target site.The coil was filled in the aneurysm and tried to detach, but the coil was unable to detach.The doctor only retracted the coil alone and switched two new coils consecutively, a galaxy g3 mini 1mm x 1cm coil (glm910010, 30889337) and a galaxy g3 4mm x 7cm (gly120407, 3162410s).However, both coils had the same issue.The doctor switched to a fourth coil to complete the surgery.The microcatheter was not replaced at all.There was no patient injury reported.Additional event information received on 14-may-2025 indicated that a pre-deployment electrical testing was performed with all involved coils.The same detachment control box (dcb) and the same cable were used successfully with the fourth coil.No neck remodeling was utilized.The coils were still attached to the coil delivery system when removed from the patient.There was no damage to the coils.The procedure was delayed by five minutes due to the event, resulting in no patient adverse consequence.
 
Additional Manufacturer Narrative
Manufacturer ref#: (b)(4).Updated sections on this medwatch: b4, d9, g3, g6, h2, h3, h6 and h11.Complaint conclusion: as reported by a health care professional, during a coil embolization to the left main coronary artery (lmca), a non-j&j microcatheter was placed and a galaxy g3 mini 2mm x 6cm coil (glm920060, 31136265) was delivered to target site.The coil was filled in the aneurysm and tried to detach, but the coil was unable to detach.The doctor only retracted the coil alone and switched two new coils consecutively, a galaxy g3 mini 1mm x 1cm coil (glm910010, 30889337) and a galaxy g3 4mm x 7cm (gly120407, 3162410s).However, both coils had the same issue.The doctor switched to a fourth coil to complete the surgery.The microcatheter was not replaced at all.There was no patient injury reported.Additional event information received on 14-may-2025 indicated that a pre-deployment electrical testing was performed with all involved coils.The same detachment control box (dcb) and the same cable were used successfully with the fourth coil.No neck remodeling was utilized.The coils were still attached to the coil delivery system when removed from the patient.There was no damage to the coils.The procedure was delayed by five minutes due to the event, resulting in no patient adverse consequence.The device was returned to j&j medtech for further evaluation.A non-sterile galaxy g3 4mm x 7cm was received contained in the decontamination pouch.Upon receiving the device, visual inspection was performed, and no appearance of damages was observed.The embolic coil was still inside the introducer.The device was inspected under microscopic magnification, and the embolic coil was found still attached to the resistance heating (rh).The distal outer sheath had not been softened, indicating that the detachment process was not initiated.Some residues of dried blood were found in the coil.The electrical resistance of the galaxy g3 device was measured with a multimeter, which is within the specification range.Also, the device was connected to a lab sample detachment control box (dcb), and to a lab sample enpower control cable, and the power was turned on.The system ready light was illuminating.Then, the detach button was pressed, and a beep sound was heard.The coil was successfully detached from the unit.The issue documented that the coil failed to detach could not be confirmed since the device met the electrical specification range.Additionally, no product defect was identified; there may have been other circumstances or issues that occurred during the use of the device that could not be replicated in the lab setting.The instructions for use (ifu) provide proper handling instructions for the device to prevent such issues from occurring.According to the risk documentation, failure to detach is a potential issue that can occur during the coil detachment due to a detachment cycle not being initiated as a result of the dpu wire detachment zone/extended coil being positioned beyond the microcatheter tip, which can result in reported failure.As part of the j&j medtech quality process, all devices are manufactured, inspected, and released to approved specifications.Therefore, no capa activity is required.A manufacturing record evaluation was performed for the finished device 3162410s number, and no non-conformances related to the malfunction were identified.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) do contain the following recommendation: verify that the microcoil delivery system is fully connected, and no faults are indicated on the dcb.If a fault exists, reseat all connections between the dpu, the dcb, and the connecting cable.If a fault still persists, replace the connecting cable.If this does not correct the error, replace the dcb.If the microcoil delivery system still continues to have a fault, retrieve the microcoil, re sheathing the microcoil system, and replace it with a new microcoil system.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY G3
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC. (FREMONT)
47709 fremont blvd
fremont CA 94538
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key22065298
Report Number3008114965-2025-00462
Device Sequence Number7849025
Product Code KRD
UDI-Device Identifier10886704080312
UDI-Public(01)10886704080312(17)271031(10)3162410S
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberGLY120407
Device Lot Number3162410S
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2025
Supplement Date Received by Manufacturer05/27/2025
Initial Report FDA Received Date05/21/2025
Supplement Report FDA Received Date06/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DETACHMENT CONTROL BOX; ENPOWER CONTROL CABLE; MICROCATHETER (UNSPECIFIED BRAND)
Patient SexUnknown
-
-