| Catalog Number |
ES2816 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Failure to Pump (1502)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/07/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
According to the available information, the device was removed and replaced as the implant was no longer pumping and there was no fluid in the reservoir.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan touch pump and cylinders 1 and 2 were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.No functional abnormalities were noted with the pump.A separation was noted on the exhaust tubing of cylinder 1 at the strain relief junction.This was a site of leakage.The surfaces appear to be rough and irregular, indicating stress was exerted.No functional abnormalities were noted with cylinder 2.Based on examination of the returned product, it was concluded that the rough and irregular surfaces associated with the separation indicates that sufficient stress was exerted on the exhaust tubing of cylinder 1 to separate the site while in-vivo.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, the device was removed and replaced as the implant was no longer pumping and there was no fluid in the reservoir.
|
| |
|
Search Alerts/Recalls
|