MAUDE Adverse Event Report: COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE Back to Search Results
Catalog Number ES2816
Medical Device Problem Codes Fluid/Blood Leak (1250); Failure to Pump (1502)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/07/2025
Type of Reportable Event Serious Injury
Event or Problem Description
According to the available information, the device was removed and replaced as the implant was no longer pumping and there was no fluid in the reservoir.
 
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
 
Additional Manufacturer Narrative
Titan touch pump and cylinders 1 and 2 were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.No functional abnormalities were noted with the pump.A separation was noted on the exhaust tubing of cylinder 1 at the strain relief junction.This was a site of leakage.The surfaces appear to be rough and irregular, indicating stress was exerted.No functional abnormalities were noted with cylinder 2.Based on examination of the returned product, it was concluded that the rough and irregular surfaces associated with the separation indicates that sufficient stress was exerted on the exhaust tubing of cylinder 1 to separate the site while in-vivo.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
 
Event or Problem Description
According to the available information, the device was removed and replaced as the implant was no longer pumping and there was no fluid in the reservoir.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Common Device Name
PENILE PROSTHESIS, INFLATABLE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key22069407
Report Number2125050-2025-00861
Device Sequence Number12929851
Product Code FHW
UDI-Device Identifier05708932539111
UDI-Public05708932539111
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/10/2025
Device Catalogue NumberES2816
Device Lot Number7776563_ES28162400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/22/2025
Supplement Report FDA Received Date09/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexMale
-
-