• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE REVERSE 42MM GLENOSPHERE; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE REVERSE 42MM GLENOSPHERE; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-01-42
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Metal Related Pathology (4530)
Date of Event 04/30/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.D10: (b)(6), - gps implant kit v2.(b)(6), 320-20-22 - eq rev compress screw lck cap kit, 4.5 x 22mm.(b)(6), 320-15-04 - rs glenoid plate post aug, 8 deg, right.(b)(6), 320-20-34 - eq rev compress screw lck cap kit, 4.5 x 34mm.(b)(6), 320-20-30 - eq rev compress screw lck cap kit, 4.5 x 30mm.(b)(6), 320-15-05 - eq rev locking screw.(b)(6), 300-01-13 - equinoxe, humeral stem primary, press fit 13mm.(b)(6), 320-20-00 - eq reverse torque defining screw kit.(b)(6), 320-20-30 - eq rev compress screw lck cap kit, 4.5 x 30mm.(b)(6), 531-20-00 - shldr gps rvrs drill kit.(b)(6), 320-10-00 - equinoxe reverse tray adapter plate tray +0.(b)(6), 531-78-20 - shouldr gps hex pins kit.(b)(6), 531-78-20 - shouldr gps hex pins kit.
 
Event or Problem Description
As reported, approximately 5 years and 1 month post the initial right reverse total shoulder arthroplasty, the patient was revised due to poly wear/fracture.The liner and glenosphere were exchanged.Photos provided indicate there may be metal-on-metal wear between the glenosphere and tray.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were corrected: h6.The revision reported was likely the result of the reported traumatic event and bone fracture.A secondary finding of edge wear of the humeral liner appears consistent with humeral liner disassociation and subsequent metal-on-metal articulation; however, these failure modes were reported to not have occurred.Additionally, the reported traumatic event and subsequent fracture could have been a contributing factor to the observed wear of the humeral liner.However, the traumatic event, fracture, and series of events cannot be confirmed and potential contributions of user or patient-related considerations to the event could not be assessed as the devices were not returned for evaluation and no radiographs or relevant clinical information were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE REVERSE 42MM GLENOSPHERE
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key22069551
Report Number1038671-2025-02118
Device Sequence Number15787298
Product Code KWT
UDI-Device Identifier10885862086396
UDI-Public10885862086396
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue Number320-01-42
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2025
Supplement Date Received by Manufacturer08/18/2025
Initial Report FDA Received Date05/22/2025
Supplement Report FDA Received Date08/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization;
Patient Age82 YR
Patient SexMale
-
-