| Catalog Number |
515005 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
Verbatim: this is a complaint about coring.It was reported that after preparing nine doses of paclitaxel, core was observed when the drug was returned to the bag.The core was small, suggesting that it originated from the injector.Additional information: could you please confirm whether coring occurred after the injector installed? if yes, could you please confirm the product code and lot number of the injector? after installing the injector, customer connected it to the protector and confirmed the occurrence when attempting to collect the liquid.The lot of the injector is unknown.Injector's product code is 515005.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) follow up mdr for device evaluation: no photos or physical samples that display the reported condition were available for investigation.Product undergoes inspections throughout the manufacturing process to ensure the quality and functionality of the device.As the lot involved in this incident is unknown, a device history review cannot be performed, and additional retained samples cannot be evaluated.Based on the available information we are not able to identify a root cause at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Search Alerts/Recalls
|