• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. WEB SL; INTRASACCULAR FLOW DISRUPTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. WEB SL; INTRASACCULAR FLOW DISRUPTION DEVICE Back to Search Results
Model Number W2-8-3
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The physical device was not available for evaluation to investigate if a condition existed that would have contributed to event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation further examine the cause of the reported event.This information may be updated if additional information is provided at a later date.This study is ongoing and device/procedure relatedness determinations can be re-adjudicated by independent reviewers/physicians.The study device and study procedure relatedness can change as a result of these reviews.Microvention is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by microvention, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Microvention has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, microvention, or its employees that the device, microvention, or its employees caused or contributed to the event described in the report.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.
 
Event or Problem Description
As reported through the web pas clinical study: target aneurysm was an unruptured basilar apex aneurysm.A 0-50% reduction in parent vessel diameter due to web impingement with no flow restriction per dsa.This event occurred after web implantation and did not require further or additional treatment/action.Patient was asymptomatic and mrs remained 0 at discharge from index procedure through to the 12 months follow up visit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEB SL
Common Device Name
INTRASACCULAR FLOW DISRUPTION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key22070952
Report Number2032493-2025-90178
Device Sequence Number9337722
Product Code OPR
UDI-Device Identifier00842429107140
UDI-Public(01)00842429107140(11)231004(17)280930(10)0000433174
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P170032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberW2-8-3
Device Lot Number0000433174
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2025
Initial Report FDA Received Date05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient SexMale
Patient Weight61 KG
-
-