Catalog Number 35251400S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/14/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation is completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "the patient was implanted with the long gamma3 nail described above on (b)(6) 2024 for a subtrochanteric fracture on the left.He was able to be mobilized under full weight-bearing during the procedure.A fracture of the long gamma nail was diagnosed during a follow-up x-ray examination, which was performed due to pain.On (b)(6) 2025, revision surgery was performed with explantation of the broken gamma 3 nail on the left.".
 
Additional Manufacturer Narrative
The reported event could be confirmed since physical inspection matches the reported failure mode; received nail was completely broken in the webs of the proximal drill hole for the lag screw.According to the damage and the evidence of the breakage surfaces, the nail broke in a fatigue fracture due to axial overload.Bearing points are typical results of the interaction of the single products under load.Areas with signs of friction / seizing [bearing points of lag screw] are visible at the lateral edge of the proximal part.Similar areas were identified at the distal medial edge.The counterparts were found on the lag screw in the areas with corresponding friction signs and worn edges.The appearance identified a high axial load application to the implants which is also supported by signs of damage verified at one locking screw.In the case presented a 67-year-old male patient was initially treated on (b)(6)2024 with above implant which was revised appr.10 months later (b)(6)2025 since nail was found to be broken during a follow-up x-ray examination, which was performed due to pain.Furthermore, it was stated per event description.He was able to be mobilized under full weight-bearing during the procedure, as neither medical records nor x-ray images of the pre and postoperative situation in different views [m-l and a-p view] were made available, a medical statement regarding initial treatment situation or the current healing status after an implant duration of appr.10 months was impossible.Nail breakage in general has been experienced but does not present an unanticipated event in itself.Depending on load application, also, depending on the patient¿s post implant behaviour and depending on suitable anatomical reduction, depending on the kind of bone breakage, depending on the course of bone healing and other factors ¿ e.G.Increased post-operative activities ¿ a nail breakage can rather be classified as anticipated; specifically, if one or more contributing issues concurrence with each other.The affected nail is designed to withstand the normal loads during the implantation period, i.E., the implant must neither be exposed to peak loads nor to continuous stresses.Another prerequisite for a successful supply is undisturbed, normal bone healing.In case that such a situation does not occur, exceeding of the fatigue strength is to be expected and thus quite predictable complications.Axial overload had led to continuous stresses and a significant weakening of the nail in the area of the lag screw thru hole, followed by the fatigue fracture.Based on the above the nail breakage was not linked to a deficiency of the device but was rather patient related.Adverse effects are clearly listed in the ifu and as pointed out ¿revision and additional surgery may be a consequence of these complications¿ a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
 
Event or Problem Description
As reported: "the patient was implanted with the long gamma3 nail described above on (b)(6)2024 for a subtrochanteric fracture on the left.He was able to be mobilized under full weight-bearing during the procedure.A fracture of the long gamma nail was diagnosed during a follow-up x-ray examination, which was performed due to pain.On (b)(6)2025, revision surgery was performed with explantation of the broken gamma 3 nail on the left.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, LEFT GAMMA3 Ø11X400MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
cécile lefeuvre
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key22072283
Report Number0008031020-2025-00910
Device Sequence Number15631160
Product Code HSB
UDI-Device Identifier07613252274315
UDI-Public(01)07613252274315(11)201014(17)250930(10)K0CA703
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K200869
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2025
Device Catalogue Number35251400S
Device Lot NumberK0CA703
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/25/2025
Supplement Date Received by Manufacturer07/17/2025
Initial Report FDA Received Date05/22/2025
Supplement Report FDA Received Date08/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
-
-