| Model Number |
DRV SFT,CANN,QC,T15 HEXALOBE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during a small surgery the screwdriver broke while inserting a screw.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Update cmackenbach 15-may-2025: further information was received that all broken parts were retrieved from the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.One unpackaged ar-8943-12, drive shaft, cannulated, qc, t15 hexalobe, serial/batch number (b)(6), was received for investigation.Upon visual inspection, it was noted that the hex tip had broken off.No fragments were returned for evaluation.No functional testing was performed due to the damage to the device.The most likely cause for the reported failure can be attributed to user-applied excessive force during the tightening process.The complaint allegation is confirmed.Refer to investigation photos.
|
| |
|
Event or Problem Description
|
|
Further information was received that a small surgery means an ankle surgery in this case.
|
| |
|