| Model Number |
B3301542M |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from a manufacturer's representative (rep) regarding an implantable neurostimulator (ins).The reason for call was to report the event.The caller indicated that they had a case today for a lead implant.During the case the surgeon bent the first lead while prepping the component. with the second lead after it was placed, they ran an impedance test with the test cable and had a value on one segment that was a little above 8000ohms.They disconnected/reconnected the lead and continued to have impedances issues a little above 8000ohms.They changed the test cable and continued to have elevated impedances so the lead was replaced.Troubleshooting was not required.The issue was not resolved through troubleshooting.
|
| |
|