| Model Number |
HX-202UR |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported during a diagnostic colonoscopy procedure, the subject device failed to deploy.The procedure was completed with an olympus back-up device and there were no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The mdr supplemental report is being submitted to provide updated fields and final investigation results.Updated fields: h2, h3, h6, h11.The device was returned for evaluation and the customer's allegation was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure.Olympus will continue to monitor the performance of this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|
Search Alerts/Recalls
|