| Model Number |
M00545150 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 49 was attempted to be used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat a common bile duct stone performed on (b)(6) 2025.During the procedure, choledochography was performed after successful insertion of the endoscope and intubation.Upon bowing the device, the cutting wire of the autotome rx 49 broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.Note: the imaging photo inside the patient shows that the cutting wire was broken.A photo of the device outside the patient also shows the cutting wire was broken and kinked.
|
| |
|
Additional Manufacturer Narrative
|
|
Block a2: patient's exact age is unknown; however, it was reported that the patient was over the age of 18.Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
|
| |
|
Additional Manufacturer Narrative
|
|
Block a2: patient's exact age is unknown; however, it was reported that the patient was over the age of 18.Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h6: imdrf device code a0406 captures the reportable event of cutting wire kinked.Block h11: (investigation result) the returned autotome rx 49 was analyzed, and a visual and media evaluation noted that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Based on the condition of the device, the wire break could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause is adverse event related to procedure.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 49 was attempted to be used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat a common bile duct stone performed on (b)(6) 2025.During the procedure, choledochography was performed after successful insertion of the endoscope and intubation.Upon bowing the device, the cutting wire of the autotome rx 49 broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.Note: the imaging photo inside the patient shows that the cutting wire was broken.A photo of the device outside the patient also shows the cutting wire was broken and kinked.
|
| |
|