| Catalog Number |
DL900J |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, an 81-year-old female patient underwent a percutaneous inferior vena cava filter implantation using a denali jugular filter via right internal jugular vein.Prior to the procedure, the filter was allegedly punctured through the sheath before it entered the patient.Another filter had to be opened.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the material puncture could not be determined based upon the provided information.Labelling/packaging review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G2, g3, h6 (method, result, conclusion).Section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, an 81-year-old female patient underwent a percutaneous inferior vena cava filter implantation using a denali jugular filter via right internal jugular vein.Prior to the procedure, the filter was allegedly punctured through the sheath before it entered the patient.Another filter had to be opened.There was no patient contact.
|
| |
|