• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FRESENIUS KABI USA LLC IVENIX, INC; LVP PRIMARY ADMINISTRATION SET, DUAL-INLET, LOW-SORBING, NEEDLE-FREE PORT, Y-SIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FRESENIUS KABI USA LLC IVENIX, INC; LVP PRIMARY ADMINISTRATION SET, DUAL-INLET, LOW-SORBING, NEEDLE-FREE PORT, Y-SIT Back to Search Results
Model Number SET-0013-01
Medical Device Problem Codes Leak/Splash (1354); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/14/2025
Type of Reportable Event Malfunction
Event or Problem Description
The following has been reported: nurse reported: tubing leak through the dial.Patient was flushed after chemotherapy, treatment was completed, tubing removed from pump then leak noted.There was no harm/exposure to patient or nurse.Nurse was wearing gloves.Not actively infusing.Verified with nurse that back priming method used during infusion was via pump function.Tubing not saved as it was used to administer hazardous drug.A preliminary review identified the following issue: administration set leak (external part) an active infusion was not stopped.Reporting as a conservative measure.No adverse effects were reported.Additional information is needed to complete the investigation.
 
Event or Problem Description
No sample or picture were available for analysis.Without the proper sample or picture evaluation it is not possible to determine the probable root cause of the reported failure.The customer complaint cannot be confirmed.The potential root causes of the leak from the cassette could be related to 1) misplaced diaphragm which resulted in a leak when the diaphragm was actuated by the valve pins in the pump, 2) an incorrect and poor sealing of the cassette in the welder machine causing the reported leakage, 3) over-insertion of the secondary port which caused a crack at the cassette, 4) cassette seal not assembled correctly causing a leak in the flow dial.The current process controls detection include 1) 100% leak and occlusion test is performed through the leak and occlusion test machine in order to capture units with any blockage or leak failure mode, 2) quality sampling for final physical testing, for performing a flow and underwater leak tests, 3) 100% visual inspection during the manufacturing process and final physical inspections.The batch record fa23l11386 was reviewed.No exceptions were generated that could classify as a possible root cause of this defect.The finished good lot has passed all sampling acceptance criteria for all tests performed including in-process testing and product testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IVENIX, INC
Common Device Name
LVP PRIMARY ADMINISTRATION SET, DUAL-INLET, LOW-SORBING, NEEDLE-FREE PORT, Y-SIT
Manufacturer (Section D)
FRESENIUS KABI USA LLC
50 high street
suite 50
north andover MA 01845
Manufacturer (Section G)
FRESENIUS KABI USA LLC
50 high street
suite 50
north andover MA 01845
Manufacturer Contact
rebecca mccandless
3 corporate drive
lake zurich, IL 60047
8475502913
MDR Report Key22077650
Report Number3014732157-2025-00578
Device Sequence Number21219338
Product Code FPA
UDI-Device Identifier00811505030054
UDI-Public00811505030054
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K183311
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberSET-0013-01
Device Catalogue NumberSET-0013-01
Device Lot NumberFA23L11386
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/23/2025
Supplement Report FDA Received Date07/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-