FRESENIUS KABI USA LLC IVENIX, INC; LVP PRIMARY ADMINISTRATION SET, DUAL-INLET, LOW-SORBING, NEEDLE-FREE PORT, Y-SIT
|
Back to Search Results |
|
| Model Number |
SET-0013-01 |
| Medical Device Problem Codes |
Leak/Splash (1354); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following has been reported: nurse reported: tubing leak through the dial.Patient was flushed after chemotherapy, treatment was completed, tubing removed from pump then leak noted.There was no harm/exposure to patient or nurse.Nurse was wearing gloves.Not actively infusing.Verified with nurse that back priming method used during infusion was via pump function.Tubing not saved as it was used to administer hazardous drug.A preliminary review identified the following issue: administration set leak (external part) an active infusion was not stopped.Reporting as a conservative measure.No adverse effects were reported.Additional information is needed to complete the investigation.
|
| |
|
Event or Problem Description
|
|
No sample or picture were available for analysis.Without the proper sample or picture evaluation it is not possible to determine the probable root cause of the reported failure.The customer complaint cannot be confirmed.The potential root causes of the leak from the cassette could be related to 1) misplaced diaphragm which resulted in a leak when the diaphragm was actuated by the valve pins in the pump, 2) an incorrect and poor sealing of the cassette in the welder machine causing the reported leakage, 3) over-insertion of the secondary port which caused a crack at the cassette, 4) cassette seal not assembled correctly causing a leak in the flow dial.The current process controls detection include 1) 100% leak and occlusion test is performed through the leak and occlusion test machine in order to capture units with any blockage or leak failure mode, 2) quality sampling for final physical testing, for performing a flow and underwater leak tests, 3) 100% visual inspection during the manufacturing process and final physical inspections.The batch record fa23l11386 was reviewed.No exceptions were generated that could classify as a possible root cause of this defect.The finished good lot has passed all sampling acceptance criteria for all tests performed including in-process testing and product testing.
|
| |
|
Search Alerts/Recalls
|
|
|