| Model Number |
126039 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: 126039 39fr rt bronco cath pu cuff x1 202104131x, 126039 39fr rt bronco cath pu cuff x1 202011095x, medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that three bronco catheters' cuffs were leaking air prior to use.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3 and h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the cuff body shows a pinhole.An attempt made to inflate the returned unit failed as the bronchial cuff could not be inflated.It was reported that three bronco catheter's cuff were leaking air prior to use.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: various bony anatomical structures (e.G., teeth) within the intubation route or any intubation tools with sharp surfaces present a threat to maintaining cuff integrity.Care must be taken to avoid damaging the thin-walled cuffs during intubation which would create the need to subject the patient to the trauma of extubation and re-intubation.If either cuff is damaged, the tube should not be used.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|