• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3543751
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 02/25/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient who underwent primary breast augmentation withs 375cc mentor memorygel breast implants experienced bilateral rupture post procedure, right sided breast pain, and right sided granuloma post procedure.The ruptures were diagnosed through mammography and magnetic resonance imagin.As a result, explant and replacement was performed on (b)(6) 2025.The right sided issues were reported initially under 1645337-2025-04439.On (b)(6) 2025 mentor received additional information and initial awareness of bilateral issues.This report relates to the left prosthesis.
 
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The investigation of the returned device was completed by the failure analysis lab on may 28, 2025.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted visual inspection and leak testing of the returned device.Visual analysis of the returned device revealed that no damage or anomalies were observed on the breast implant.Leak testing was performed, in accordance with mentor procedures, and no areas of gel exposure were detected during the analysis.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture status identified by mri evaluation includes both suspected ruptures, which are those ruptures identified by mri but not confirmed by explantation and examination of the device, and confirmed ruptures, which are those ruptures that are confirmed by evaluation of the explanted devices.The event described could not be confirmed as the breast implant was returned without detectable gel exposure.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key22078597
Report Number1645337-2025-05715
Device Sequence Number18476063
Product Code FTR
UDI-Device Identifier00081317000891
UDI-Public(01)00081317000891(17)210228(10)7290221
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number3543751
Device Lot Number7290221
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2025
Supplement Date Received by Manufacturer05/28/2025
Initial Report FDA Received Date05/23/2025
Supplement Report FDA Received Date05/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-