Model Number K-LESS T-ROPE W/DRV, SYN REPR, SS
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a syndesmosis surgery when the surgeron tried to release the button of the tightrope, it did not flip.As he pulled it, the button got stuck inside the patients bone.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a different device ar-8926ss.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or improper bone preparation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
K-LESS T-ROPE W/DRV, SYN REPR, SS
Common Device Name
WASHER, BOLT, NUT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22078990
Report Number1220246-2025-02184
Device Sequence Number8530825
Product Code HTN
UDI-Device Identifier00888867281615
UDI-Public(01)00888867281615(11)240911(17)290831(10)15315602
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K043248
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberK-LESS T-ROPE W/DRV, SYN REPR, SS
Device Catalogue NumberAR-8925SS
Device Lot Number15315602
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2025
Supplement Date Received by Manufacturer10/28/2025
Initial Report FDA Received Date05/23/2025
Supplement Report FDA Received Date10/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-