It was reported that the patient controller exhibited a controller fault alarm indicating the internally battery end of life.The patient came into the clinic, and autologs reported two controller fault alarms.A successful controller exchange was performed in the intensive care unit, with the clinician exchanging the old controller for a new primary controller, and the ventricular assist device (vad) started right up.The old controller was disposed of by the clinician and will not be returned for analysis.No patient complications have been reported as a result of this event.
|
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
A supplemental report is being submitted for analysis and investigation completion.Product event summary: the controller ((b)(6)) was not returned for evaluation as it was disposed of.A review of the device history record was not performed as the reported event and analysis are not related to a manufacturing or servicing issue.Review of the controller log files revealed two (2) controller fault alarms logged on (b)(6) 2025 at 06:35:14 and at 09:12:58, as well as multiple event exceptions logged on (b)(6) 2025, indicating an issue with the internal battery.In addition, log files revealed that the controller had been in use for more than two (2) years.As a result, the reported event was confirmed.Applicable risk documentation, experience with events of similar circumstances, and the available information were considered; possible root causes of the reported controller fault alarm event may be attributed, but not limited, to a reduced charge capacity of the internal battery and/or a faulty internal battery charger integrated circuit.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|