ETHICON INC. PRLNE BLU 24IN 7-0 D/A CC175-8 MP; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
|
Back to Search Results |
|
| Catalog Number |
W8704 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H6.Component code: g07002 - device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: was surgery delayed due to the reported event? no, was procedure successfully completed? yes, were fragments generated? no, if yes, were they removed easily without additional intervention? unknown, patient status/ outcome / consequences: no, the following information was requested, but unavailable: was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study: unknown, attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.1.How many devices demonstrated the alleged deficiency (initial report mention 9 devices involved)? 2.Please specify if the suture broken or the needle pull off of the suture? this report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an av canal repair procedure on an unknown date and suture was used.When the surgeon used the device*2ea to suture, this suture was broken.The breakpoint in the junction between needle and suture.Then, he used other lot number to complete this procedure.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6, d9, h3.H3 investigation summary: the product was returned to ethicon for evaluation.One empty overwrap packet that pertains to product code w8704 was received for analysis.The visual assessment of the product noted that since the sutures and needles were not returned, the event described could not be confirmed.If additional information becomes available, the investigation will be updated accordingly.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.No conclusion could be reached on the cause of the reported event.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional h3 investigation summary: the product was returned to ethicon for evaluation.One open box with eight unopened samples were received for analysis.Product code w8704.In order to evaluate the condition of the returned samples, the packets were opened.The swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along the strand and no issues related to breakage sutures or anomalies were observed during the evaluation.The functional test was performed using instron equipment and the tensile force was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch and no non-conformances were identified.Additional information was requested, and the following was obtained: 1.How many devices demonstrated the alleged deficiency (initial report mention 9 devices involved)? >> hcp experienced 2 ea with suture broken, then have the concern for the rest of 7ea unopen suture with the similar issue.2.Please specify if the suture broken or the needle pull off of the suture? >> suture broken.3.Please provide the title of the external person providing answers to follow-up (external person submitting answers to sales rep) >> operating room nurse.
|
| |
|
Search Alerts/Recalls
|
|
|