| Catalog Number |
80326 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.H11: d2b (dqx; qew).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent a thrombectomy and atherectomy procedure via left lower arm using aspirex.During the procedure, the hand switch allegedly came apart and the breakage occur outside of the patient body.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H.11.: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: no physical sample was received.A physical investigation was not possible.The user report and provided image contains information regarding catheter detachment.After review of the provided images detachment can be confirmed.A clear root cause for these incidents could not be identified but break of the guidewire represents a known inherent risk.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D2b (dqx; qew), g2, g3, h6 (component, method, result, conclusion).Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent a thrombectomy and atherectomy procedure via left lower arm using aspirex.During the procedure, the hand switch allegedly came apart and the breakage occur outside of the patient body.The procedure was completed using another device.There was no reported patient injury.
|
| |
|