|
It was reported the thunderbeat surgical instrument¿ s blade head fractured during a therapeutic laparoscopic procedure.The fractured parts did not fall within the patient anatomy.The procedure was completed using a backup olympus device.There was no delay, patient injury, nor medical intervention associated with this event.
|
|
This supplemental report is being submitted to provide the legal manufacturer¿s final investigation.Additional data:h3 the device was returned to olympus for inspection, and the reportable malfunction was confirmed.Based on the results of the investigation, a definitive root cause could not be established.It is likely the following led to the malfunction: 1.During output in seal & cut mode, the probe came in contact with hard tissue, metal or a surgical instrument.Consequently, a scratch was made on the probe.2.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch.3.The probe broke when added load.The event can be detected/prevented by following the applicable chapters in the instructions for use which state: ¿ the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator, forceps, and others).Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/splitting/protruding/partial separating of the tissue pad.In turn, the probe may break before displaying an error window or generating an alarm tone.¿ do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.Olympus will continue to monitor field performance for this device.
|