| Model Number |
GIF-XP170N |
| Medical Device Problem Code |
Melted (1385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: it is possible that a high-energy treatment device (such as a high-frequency device) was used without ensuring a sufficient distance from the tip.The event 2 is detectable and preventable by handling device in accordance with the following ifu.Gastrointestinal videoscope gif-xp170n operation manual chapter 3 preparation and inspection: 3.3 inspection of the endoscope: inspection of the endoscope gastrointestinal videoscope gif-xp170n operation manual chapter 4 operation: 4.3 using endotherapy accessories.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the gastrointestinal videoscope exhibited a burn at the distal end.There was no patient involvement.
|
| |
|