| Model Number |
SIG30AVM |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: egiaushort - egiaushort endogia ultra univ sht stap - lot# p4h0843.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a open appendectomy resection, after clamping the tissue, the handle for the firing could not be squeezed.There were no problems with the opening and closing of the jaw; however, the jaw was closed contrary to the intention from the beginning.The handle was replaced at first, but it could not be clamped due to the same event, so the cartridge was replaced and the firing could be done.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d3 (mfr name and address), d4 (udi#), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found that the reload was loaded into a representative pmv handle.The interlock was overridden and the reload was applied to test media.All staples were placed, and test media was transected.The reload interlock was tested and found to function properly.It was reported that the handle for the firing could not be squeezed.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the firing button had been partially compressed and then the open button was pressed after pressing the green firing button.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to completely fire the reload will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis and/or leakage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|