BARD PERIPHERAL VASCULAR, INC. SENOMARK ULTRA BREAST TISSUE MARKER; BIOPSY INSTRUMENT Back to Search Results
Catalog Number SMEC10R
Medical Device Problem Codes Positioning Failure (1158); Difficult to Insert (1316); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On april 24, 2025, a patient underwent an electron-triggered x-ray imaging breast tissue marker placement procedure through grease density tissue using senomark ultra and encor probe.During the procedure, the marker was allegedly difficult to insert into the probe.It was further reported that the marker failed to deploy.After washing and placing the marker on the needle, the marker did not come out of the guide, remaining halfway open, which made it difficult to remove the needle from the breast.Reportedly needle was removed with the help of an emergency surgeon, by making a larger incision.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the alleged difficult to insert, positioning failure and difficult to remove issue could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (unique identifier (udi) #), g3, h6 (component, method).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2025, a patient underwent an electron-triggered x-ray imaging breast tissue marker placement procedure through grease density tissue using senomark ultra and encor probe.During the procedure, the marker was allegedly difficult to insert into the probe.It was further reported that the marker failed to deploy.After washing and placing the marker on the needle, the marker did not come out of the guide, remaining halfway open, which made it difficult to remove the needle from the breast.Reportedly needle was removed with the help of an emergency surgeon, by making a larger incision.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENOMARK ULTRA BREAST TISSUE MARKER
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
INFUS MEDICAL (THAILAND)
706 moo 4
bangpoo ind estate
samutprakarn province 10280
TH   10280
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key22083139
Report Number2020394-2025-00792
Device Sequence Number8060383
Product Code NEU
UDI-Device Identifier00801741032035
UDI-Public(01)00801741032035(11)230428(17)260401(10)VTHR00204
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K123911
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2026
Device Catalogue NumberSMEC10R
Device Lot NumberVTHR00204
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2025
Supplement Date Received by Manufacturer08/29/2025
Initial Report FDA Received Date05/26/2025
Supplement Report FDA Received Date09/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age45 YR
Patient SexFemale
Patient Weight64 KG
-
-