|
SYNTHES GMBH WASHER FOR SCREW 14/7MM STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
04.045.780S |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H111: additional narrative: d4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was implanted with a retrograde femoral nail advanced (rfna) on (b)(6) 2025.An open reduction and retrograde intramedullary nailing of the left femur surgery occurred on (b)(6) 2025.Post-operatively x-rays showed significant back-out and the screw was prominent.The patient had limited fracture healing at that point, so the patient underwent a revision procedure on (b)(6) 2025, for screw exchange and placement of a locking nut that threads onto the screw.Later it was discovered that the locking nut and screw were disengaged, and the screw was partially backed out.On (b)(6) 2025, another surgery was performed for left femur removal of hardware, revision of im nailing.Post-operative follow-up indicate that the patient appears stable, and imaging showed no loosening or failures at the present time; the fracture alignment remains stable.This report captures the third removal event that occurred on (b)(6) 2025.The first removal event that occurred on (b)(6) 2025, is captured on related complaint (b)(4).The second removal event that occurred on (b)(6) 2025, is captured on related complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d10 h6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device (part # 04.045.780s-us lot # 2476p01) and no non-conformances / manufacturing irregularities were identified.If additional information is made available, the investigation will be updated as applicable.
|
| |
|
Search Alerts/Recalls
|
|
|