• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH LOCKING SCREW FOR IM NAIL 5/ 86/ XL25; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH LOCKING SCREW FOR IM NAIL 5/ 86/ XL25; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 04.045.086
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was implanted with a retrograde femoral nail advanced (rfna) on (b)(6) 2025.An open reduction and retrograde intramedullary nailing of the left femur surgery occurred on (b)(6) 2025.Post-operatively x-rays showed significant back-out and the screw was prominent.The patient had limited fracture healing at that point, so the patient underwent a revision procedure on (b)(6) 2025, for screw exchange and placement of a locking nut that threads onto the screw.Later it was discovered that the locking nut and screw were disengaged, and the screw was partially backed out.On (b)(6) 2025, another surgery was performed for left femur removal of hardware, revision of im nailing.Post-operative follow-up indicate that the patient appears stable, and imaging showed no loosening or failures at the present time; the fracture alignment remains stable.This report captures the third removal event that occurred on (b)(6) 2025.The first removal event that occurred on (b)(6) 2025, is captured on related complaint (b)(4).The second removal event that occurred on (b)(6) 2025, is captured on related complaint (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d10 h6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.Device history review: the combination of lot nr.26124p2 and part nr.04.045.086 doesn't exist.The correct part.Nr for this lot should be 04.045.086.Please confirm if additional information is made available, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW FOR IM NAIL 5/ 86/ XL25
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key22084173
Report Number8030965-2025-05196
Device Sequence Number6947784
Product Code HWC
UDI-Device Identifier10886982293381
UDI-Public(01)10886982293381(10)26124P2
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K201346
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number04.045.086
Device Lot Number26124P2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/30/2025
Supplement Date Received by Manufacturer06/11/2025
Initial Report FDA Received Date05/26/2025
Supplement Report FDA Received Date06/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LOCKING SCREW FOR IM NAIL 5/ 38/ XL25; LOCKING SCREW FOR IM NAIL 5/ 38/ XL25; LOCKING SCREW FOR IM NAIL 5/ 60/ XL25; RFNA/ 11MM/ 360MM/ STANDARD BEND/ STER; RFNA/ 11MM/ 360MM/ STANDARD BEND/ STER; UNK - NUTS; WASHER FOR SCREW 14/7MM STERILE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-