ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that during an intraocular lens (iol) implant procedure, the lens was inserted and they noticed a crack in the lens.So, it was then removed and a backup lens was inserted.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information was provided in h.3.,h.6 and h.11.The product was not returned for analysis.The complainant indicates the use of non-company viscoelastic which is not qualified for use with the associated intra ocular lens (iol) model.The reported complaint was not observed as no sample was returned for analysis, however, based on the information provided by the customer, the most likely root cause is failure to follow instructions for use (ifu), as the surgeon states the use of non-qualified viscoelastic.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22084659
Report Number9612169-2025-01115
Device Sequence Number6736103
Product Code HQL
UDI-Device Identifier00380655093139
UDI-Public(01)00380655093139(17)270822(11)220824(10)25478734
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.170
Device Lot Number25478734
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/30/2025
Supplement Date Received by Manufacturer07/15/2025
Initial Report FDA Received Date05/26/2025
Supplement Report FDA Received Date08/05/2025
Date Device Manufactured08/24/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMVISC PLUS (LOT 030733); B&L 702510; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Patient SexUnknown
-
-