| Model Number |
SN60WF |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A nurse reported that during an intraocular lens (iol) implant procedure, the lens was inserted and they noticed a crack in the lens.So, it was then removed and a backup lens was inserted.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3.,h.6 and h.11.The product was not returned for analysis.The complainant indicates the use of non-company viscoelastic which is not qualified for use with the associated intra ocular lens (iol) model.The reported complaint was not observed as no sample was returned for analysis, however, based on the information provided by the customer, the most likely root cause is failure to follow instructions for use (ifu), as the surgeon states the use of non-qualified viscoelastic.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|