MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT Back to Search Results
Catalog Number 788626
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/16/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during ureteral stenosis retrograde pyelogram, it was confirmed that the tip of the tiger tail catheter had come off.The tip was confirmed by fluoroscopy.It was removed.Per additional information received on 20may2025, stated that the tip of the tiger tail ureteral catheter was severed.Fluoroscopy with rp (ureteral contrast examination) confirmed that the tip had come off and remained inside the body.Observed with a ureteroscope.Removed with relic forceps.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The reported event was confirmed ¿ supplier related.Sample has been evaluated and observed the black tip of guidewire was detached.A potential root cause for this failure mode could be, ¿defect components from supplier¿.A review of the device history record did not show any problems or condition that would have contributed to the reported issue.The instructions for use were found adequate and states the following: the inlay optima ureteral stent and multi-length ureteral stent with suture are indicated to relieve obstruction in a variety of benign, malignant and post-traumatic conditions in the ureter.These conditions include stones and/or stone fragments, or other ureteral obstructions such as or ureteral trauma, or in association with extracorporeal shock wave lithotripsy (eswl).The stent may be placed using endoscopic surgical techniques or percutaneously using standard radiographic technique.It is recommended that the indwelling time not exceed 365 days.The stent is not intended as a permanent indwelling device.Description: the inlay optima ureteral stent and multi-length ureteral stent is a coated, double pigtail available in two forms: a single size or a customizable multi-length size.The following items are included with each stent: 1 ureteral stent with suture.1 push catheter with radiopaque band.1 pigtail straightener.1 guidewire (optional).Note: a 4.7 fr stent is compatible with a.035 guidewire and 6, 7, and 8 fr stents are compatible with a.038 guidewire.In vitro testing conducted on the inlay optima ureteral stent and multi-length ureteral stent indicate reduced accumulation of urine calcium salts as assayed by calcium when compared to a control.Correlation of in vitro data to clinical outcome has not been established.Urethral catheters, and polymers intended for urological use, bju international (2000), 86, 414-421.Contraindications: no known contraindications for use.Precautions: suture may be cut off prior to stent placement.Remove suture if indwelling time is expected to be longer than 14 days.The overall integrity of the stent.Ureteral stents should be checked periodically for signs of encrustation and proper function.Periodic checks of the stent by cystoscopic and/or radiographic procedures are recommended at intervals deemed to be appropriate by the physician when long-term use is indicated, it is recommended that indwelling time not exceed 365 days.The stent is not intended as a permanent indwelling device.With any ureteral stent, migration is a possible complication, which could require medical intervention for removal.Selection of too short a stent may result in migration.Care should be exercised when removing the stent from the inner polybag to eliminate tearing or fragmentation.The insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure.Multi-length ureteral stents: formation of knots in multi-length ureteral stents may occur.This may result in injury to the ureter during removal and/or the need for additional surgical encountered during attempts at removal.Potential complications associated with retrograde/antegrade positioning of indwelling ureteral stents include the following: edema, stone formation, peritonitis fistula formation, loss of renal function fragmentation, migration, occlusion hemorrhage, pain/discomfort, stent encrustation hydronephrosis, perforation of kidney, renal, ureteral erosion infection pelvis, ureter and/or bladder, urinary symptoms directions for use: 1.Determine the proper stent length for the patient.This is generally calculated from the baseline 2.Insert the cystoscope then pass the guidewire through the scope until the tip is in the renal pelvis.3.Move the pigtail straightener over the proximal end (kidney coil end) of the ureteral stent allowing easier insertion onto the guidewire.Remove pigtail straightener once the stent is secure on the guidewire.4.Pass the stent over the guidewire through the cystoscope by using the push catheter for proper placement.5.Watch the distal end (bladder coil end) of the stent or the radiopaque, proximal end of the pusher.Stop advancing when the stent¿s distal end marker reaches the ureterovesical junction (uvj).(see below for proper placement directions on the multi-length ureteral stent.) 6.Withdraw the guidewire slowly.The stent will form a pigtail automatically.7.Carefully remove the push catheter.Packaging.Multi-length ureteral stent placement: to accurately size this stent count the marker bands second and third bands indicate 24cm and 26cm lengths respectively.The last large band is the 28cm length.If you need to place for the 30cm and 32cm lengths, use the attached suture or endoscopic forceps to gently pull back on the stent unwinding the coil from the kidney.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and with applicable local, state and federal laws and regulations.This is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Correction- d, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during ureteral stenosis retrograde pyelogram, it was confirmed that the tip of the tiger tail catheter had come off.The tip was confirmed by fluoroscopy.It was removed.Per additional information received on 20may2025, stated that the tip of the tiger tail ureteral catheter was severed.Fluoroscopy with rp (ureteral contrast examination) confirmed that the tip had come off and remained inside the body.Observed with a ureteroscope.Removed with relic forceps.
 
Event or Problem Description
It was reported that during ureteral stenosis retrograde pyelogram, it was confirmed that the tip of the tiger tail catheter had come off.The tip was confirmed by fluoroscopy.It was removed.Per additional information received on 20may2025, stated that the tip of the tiger tail ureteral catheter was severed.Fluoroscopy with rp (ureteral contrast examination) confirmed that the tip had come off and remained inside the body.Observed with a ureteroscope.Removed with relic forceps.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Correction- d, g this report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during ureteral stenosis retrograde pyelogram, it was confirmed that the tip of the ureteral stent had come off.The tip was confirmed by fluoroscopy.It was removed.Per additional information received on 20may2025, stated that the tip stent was severed.Fluoroscopy with rp (ureteral contrast examination) confirmed that the tip had come off and remained inside the body.Observed with a ureteroscope.Removed with relic forceps.
 
Additional Manufacturer Narrative
The reported event was confirmed ¿ cause unknown.The product had caused the reported failure.Sample has been evaluated and observed the black tip of tigertail had detached.Communication was sent to suppliers to address this issue.Based on the supplier¿s response, this complaint is unknown cause as their inspection records indicated no abnormalities.During the sub-assembly manufacturing process at the supplier¿s facility, all production lots underwent 100% visual inspection.This included flexing the tip at a 90° angle under a microscope to detect any abnormalities.Additionally, pull test measurements were recorded for the first and every 21st part in each batch.Based on the inspection records, no abnormalities were observed for the affected component or batch.A potential root cause for this failure mode could be due to user related or supplier.The labeling and instructions for use were found to be adequate.A review of the device history record did not show any problems or condition that would have contributed to the reported issue.Therefore, no additional action required at this time.Corrections: e, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY OPTIMA® URETERAL STENT
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key22088212
Report Number1018233-2025-03860
Device Sequence Number7885481
Product Code FAD
UDI-Device Identifier00801741015762
UDI-Public(01)00801741015762(11)250204(17)271228(10)MYJZ0998
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K043193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number788626
Device Lot NumberMYJZ0998
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2025
Supplement Date Received by Manufacturer06/13/2025
07/22/2025
11/26/2025
Initial Report FDA Received Date05/27/2025
Supplement Report FDA Received Date07/01/2025
07/24/2025
12/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-