| Model Number |
0184 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited a gradual rise in shock lead impedance measurements and high out of range pace impedance measurements.Technical services (ts) noted that this was likely associated with spring contact and calcification.It was also noted that there was no noise on stored events and no changes to threshold.Ts recommended performing a shock vector breakdown and programming changes.This rv lead remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific and physical analysis could not be conducted.Investigation of the available information/data determined that the observed gradual rise in lvsi measurements was likely associated with calcification, as there was no conclusive evidence of compromised physical or electrical integrity of the lead.This condition may develop over time and can be reflected in a gradual and sustained increase in shock impedance measurements.Boston scientific issued a field safety notice to physicians in july 2025 regarding the potential for reliance defibrillation leads with expanded polytetrafluoroethylene (eptfe) coating to exhibit a pattern of gradually increasing lvsi measurements due to calcification, which may result in elevated high voltage shock impedance (hvsi) measurements and reduced shock efficacy.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited a gradual rise in shock lead impedance measurements and high out of range pace impedance measurements.Technical services (ts) noted that this was likely associated with spring contact and calcification.It was also noted that there was no noise on stored events and no changes to threshold.Ts recommended performing a shock vector breakdown and programming changes.This rv lead remains in service.No adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|