• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY? XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY? XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941824350
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/23/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that shaft break and balloon detachment occurred resulting in surgical intervention.The target lesions were located in the left main coronary artery (lmca), left anterior descending artery (lad), and left circumflex artery (lcx).After a 7f non-boston scientific (bsc) sheath was inserted within the existing impella sheath using a single-access technique, a 7f non-bsc guide catheter was advanced in the ostium of lmca.Then, two separate non-bsc guidewires were advanced into the lcx and lad, and pre-dilated the lesions with 2.5mm nc emerge balloon catheters.Following pre-dilation, an intravascular lithotripsy was performed using a 3.0 x 12mm shockwave balloon to treat the lmca and lad.Subsequently, a 3.5x28mm synergy drug-eluting stent (des) was deployed in the ostial lmca to proximal lad, followed by 3.50 x 24mm synergy xd des in the ostial lmca to proximal lcx.These were post-dilated at high pressure and finished with simultaneous kissing balloon inflation to achieve a final lmca diameter of 6 mm.The final result showed excellent blood flow (timi 3), with no residual narrowing, dissections, blockages, or complications, and the patient was not suitable for atherectomy.However, during removal of the stent delivery system from the lcx, the shaft broke, separated the balloon from the main shaft and left inside the patient.Attempts were made to retrieve the device with a snare but was unsuccessful.The case was discussed, and to avoid potentially dislodging the implanted stent or causing vessel injuries, the patient was immediately transferred to a tertiary center for a coronary artery bypass graft.All wires and catheters were removed at the end, and the impella and sheaths were secured in place.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY? XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key22089539
Report Number2124215-2025-31177
Device Sequence Number6947607
Product Code NIQ
UDI-Device Identifier08714729981060
UDI-Public(01)08714729981060(17)260827(10)34802180
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/27/2026
Device Model NumberH7493941824350
Device Catalogue NumberH7493941824350
Device Lot Number0034802180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2025
Initial Report FDA Received Date05/27/2025
Date Device Manufactured08/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
-
-