• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC UNIVERSA SOFT URETERAL STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNIVERSA SOFT URETERAL STENT SET; FAD STENT, URETERAL Back to Search Results
Model Number G49937
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.E1 - facility name: (b)(6) public and private hospitals.E1 - phone number: (b)(6).H3: device evaluation is anticipated but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported a stent from a universa soft ureteral stent set was found broken at the tip around 1 cm with the tethers still on it, prior to an unspecified procedure.The device was not inserted into the patient and the issue was discovered when the user was preparing it for implant.The procedure was completed by using another same-type device.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation.It was reported a stent from a universa soft ureteral stent set was found broken at the tip around 1 cm with the tethers still on it, prior to an unspecified procedure.The device was not inserted into the patient and the issue was discovered when the user was preparing it for implant.The procedure was completed by using another same-type device.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturer¿s instructions (mi), and quality control (qc) procedures were conducted during the investigation.Functional tests and visual inspection of the returned complaint device was also conducted.One device was returned for investigation in an open package with label.The tip was visually confirmed to be separated from the stent.The portion of the tip that was separate had the tether still attached to it.The point of separation was at a sideport.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no related non-conformances reported for the failure mode.A complaint history database search showed no other related complaints associated with the failure mode for the complaint device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.Cook also reviewed product labeling; the ifu supplied with the device states the following in consideration of the reported failure mode: how supplied "upon removal from the package, inspect the product to ensure no damage has occurred." review of the device history record, complaint history, quality control documents, and device failure analysis indicates that the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Based upon the available information, inspection of the returned device, and results of the investigation, cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSA SOFT URETERAL STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key22089916
Report Number1820334-2025-00596
Device Sequence Number18148261
Product Code FAD
UDI-Device Identifier00827002499379
UDI-Public(01)00827002499379(17)280122(10)16431424
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K151051
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG49937
Device Catalogue NumberUSH-500
Device Lot Number16431424
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2025
Supplement Date Received by Manufacturer08/20/2025
Initial Report FDA Received Date05/27/2025
Supplement Report FDA Received Date08/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-