Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation.It was reported a stent from a universa soft ureteral stent set was found broken at the tip around 1 cm with the tethers still on it, prior to an unspecified procedure.The device was not inserted into the patient and the issue was discovered when the user was preparing it for implant.The procedure was completed by using another same-type device.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturer¿s instructions (mi), and quality control (qc) procedures were conducted during the investigation.Functional tests and visual inspection of the returned complaint device was also conducted.One device was returned for investigation in an open package with label.The tip was visually confirmed to be separated from the stent.The portion of the tip that was separate had the tether still attached to it.The point of separation was at a sideport.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no related non-conformances reported for the failure mode.A complaint history database search showed no other related complaints associated with the failure mode for the complaint device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.Cook also reviewed product labeling; the ifu supplied with the device states the following in consideration of the reported failure mode: how supplied "upon removal from the package, inspect the product to ensure no damage has occurred." review of the device history record, complaint history, quality control documents, and device failure analysis indicates that the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Based upon the available information, inspection of the returned device, and results of the investigation, cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|