• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1600-05
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Codes Failure of Implant (1924); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Event or Problem Description
The recipient is reportedly experiencing decreased performance and sound quality issues.A review of the recipient's test data indicates impedance issues.Revision surgery is scheduled.
 
Additional Manufacturer Narrative
Additional information: sections b.3, d.6b, h.6.The recipient's device was explanted.The recipient was re-implanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed and silicone damage was observed on the top and bottom cover of the device.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was obtained.The condition of the electrode prevented an electrical test from being performed.The device passed some of the electrical tests performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
Additional information: section d.9.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA IMPLANT
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
heidi white
28515 westinghouse place
valencia, CA 91355
MDR Report Key22090055
Report Number3006556115-2025-00669
Device Sequence Number15137454
Product Code MCM
UDI-Device Identifier07630016855696
UDI-Public(01)07630016855696(11)181018(17)211031
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
P960058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/31/2021
Device Model NumberCI-1600-05
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Not provided
Initial Report FDA Received Date05/27/2025
Supplement Report FDA Received Date07/21/2025
09/29/2025
10/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/18/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age7 YR
Patient SexFemale
-
-