This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, h2, h3, h6, h11.The device was returned to olympus for inspection and the reported failure was not confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: gas leak in secondary pipeline due to swollen tube a root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: regarding the customer's allegation, no problem was detected.Regarding the reportable issue found during the investigation, the cause is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|