BOSTON SCIENTIFIC CORPORATION LATITUDE PROGRAMMING SYSTEM; EXTERNAL PROGRAMMER Back to Search Results
Model Number 3300
Medical Device Problem Codes Pacing Problem (1439); Defective Device (2588); Device Sensing Problem (2917); Key or Button Unresponsive/not Working (4063)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the pacing system analyzer (psa) on this integrated programmer was unable to pace or sense in neither of the port.Additionally, left ventricle port pace but noise was noted and screen froze.Technical services (ts) suggested to have the programmer be returned.No patient involvement reported.
 
Additional Manufacturer Narrative
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory, including visual inspection, functional and baseline testing, and review of device memory.Analysis confirmed the programmer touchscreen was unresponsive due to not detecting touch inputs.Broken traces were noted on a flexible cable within the touchscreen assembly, which resulted in the clinical observation of an unresponsive touchscreen.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.Investigation determined that the most common cause of the touchscreen becoming unresponsive during use was the water detection feature within the programmer's liquid crystal display (lcd) touchscreen assembly.The purpose of this water detection feature is to detect the presence of water on/near the programmer screen.If water is detected, this feature is designed to disable the touchscreen to prevent water droplets from causing false positive touch inputs during use.Specifically, the investigation determined that various inputs could induce false positive detections of this water detection feature, which then results in an unresponsive touchscreen.Boston scientific's investigation also identified additional potential causes independent of the water detection feature.These include a variety of unrelated root causes such as hardware, environmental, software, and user-related causes that can be intermittent and not reproducible during laboratory analysis.
 
Event or Problem Description
It was reported that the pacing system analyzer (psa) on this integrated programmer was unable to pace or sense in neither of the port.Additionally, left ventricle port pace but noise was noted and screen froze.Technical services (ts) suggested to have the programmer be returned.No patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATITUDE PROGRAMMING SYSTEM
Common Device Name
EXTERNAL PROGRAMMER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
PLEXUS ELECTRONICA S DE RL DE CV
paseo del norte no
4640 guadalajara
zapopan, jal 45010
MX   45010
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22090574
Report Number2124215-2025-34339
Device Sequence Number6926213
Product Code OSR
UDI-Device Identifier00802526620201
UDI-Public(01)00802526620201
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P910077/S159
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number3300
Device Catalogue Number3300
Device Lot Number027923
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2025
Supplement Date Received by Manufacturer10/10/2025
Initial Report FDA Received Date05/27/2025
Supplement Report FDA Received Date10/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1