Correction to h6(results code).The complaint couldn't be confirmed, since the returned image for evaluation don't matches the alleged failure.Upon further investigation of the ct scans by healthcare professionals the following was observed: the ct scan does not show any abnormalities regarding the tibial component.There is a subsidence described for the talar component.Adjacent to the talar component there is almost no radiolucence, and no large cysts, the device, however, seems to be subsided medially a little more, than laterally.Without the comparison to initial radiographs or ct scan after the implantation, this is difficult to assess.The underlying cause of the subsidence is unclear; it could be related to poor bone substance and thus a patient related factor.However, failure of total ankle replacement tar over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient radiological and clinical information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided, because key clinical information e.G., clinical status, exact symptoms, and range of motion of the patient, assessment of the treating physician is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitations, we are unable in such cases to provide statements about the patient, the procedure and or the device in relation to cause of the failure.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|