| Catalog Number |
10014914 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bd alaris syringe administration set was occluded.The following information was received by the initial reporter with the following verbatim: it was reported by customer that when running in the syringe pump, at appropriate rates, they are beeping ¿occluded¿ and then when inspected and hand flushed, they begin spraying out the side of the connection to patient or spraying out of the disc.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
No product or photo was returned by the customer.The customer complaint of flow issues: fluid blockage could not be verified due to the product not being returned for failure investigation.A device history record review for material # 10014914 and lot # 24125244 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 13dec2024.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|