• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INBONE TIBIAL TRAY LEFT SIZE 3 LEFT; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INBONE TIBIAL TRAY LEFT SIZE 3 LEFT; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 200252903
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Post Operative Wound Infection (2446); Insufficient Information (4580)
Date of Event 05/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient underwent a device explant surgery for reasons that are not available at the time of this report.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient underwent a device explant surgery due to infection.The patient received a cement spacer.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Microbiologist reviewed the sterilization procedures, environmental monitoring, bioburden data and the dhr and noted: the subject device was packaged according to established design and process specifications and was sterilized according to procedure.No deviation for a non-conformance could be found.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INBONE TIBIAL TRAY LEFT SIZE 3 LEFT
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key22092593
Report Number3010667733-2025-00660
Device Sequence Number18109556
Product Code HSN
UDI-Device Identifier00840420124142
UDI-Public(01)00840420124142(11)230918(17)310918(10)1742040
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K051023
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number200252903
Device Lot Number1742040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/01/2025
Supplement Date Received by Manufacturer09/18/2025
Initial Report FDA Received Date05/27/2025
Supplement Report FDA Received Date10/16/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-