| Model Number |
A22202C |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: (b)(6) hospital.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the electrode loop broke on the probe.The issue occurred during a transurethral resection of the prostate procedure and was completed using a similar device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: no additional information received from customer.D9: additional information g1: correction h2: additional information, device evaluation, correction h3: additional information h4: additional information h6: additional information the device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: there were charred marks on the distal end side.A root cause for the reported events could not be identified.Based on the results of the investigation, the most likely cause was also not established.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|