• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM? CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM? CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) Back to Search Results
Model Number G126
Medical Device Problem Codes Signal Artifact/Noise (1036); Failure to Capture (1081)
Health Effect - Clinical Code Device Overstimulation of Tissue (1991)
Date of Event 04/24/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system exhibited noise and suspected loss of capture (loc) on the left ventricular (lv) lead channel.The health care professional performed an in-office evaluation and noted changes in morphology during isometrics as well as phrenic nerve stimulation in certain pacing vectors.No adverse patient effects were reported.This system remains in service.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system exhibited noise and suspected loss of capture (loc) on the left ventricular (lv) lead channel.The health care professional performed an in-office evaluation and noted changes in morphology during isometrics as well as phrenic nerve stimulation in certain pacing vectors.No adverse patient effects were reported.This system remains in service.Additional information was received indicating that the physician opted to turn the left ventricular (lv) lead pacing off and allow intrinsic activity to occur.No additional adverse patient effects were reported.This system remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM? CRT-D
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22093226
Report Number2124215-2025-34560
Device Sequence Number17879420
Product Code NIK
UDI-Device Identifier00802526589140
UDI-Public(01)00802526589140(17)231213
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/13/2023
Device Model NumberG126
Device Catalogue NumberG126
Device Lot Number106163
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2025
Supplement Date Received by Manufacturer05/30/2025
Initial Report FDA Received Date05/27/2025
Supplement Report FDA Received Date06/02/2025
Date Device Manufactured01/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age79 YR
Patient SexMale
-
-