D10 concomitant products: vlocm0804, vlocm0804 v-loc* 90 3-0 vio 15cm cv23, (lot #a4f1959vy) vlocm0804, vlocm0804 v-loc* 90 3-0 vio 15cm cv23, (lot #a4f1959vy) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Additional information: d3 (mfr name and street 1), d4 (udi), d9, g3, h3, h6.Correction: a5a, a5b, e1 (facility name, street 1, city, region, country, postal code, fax number and phone number).H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted one needle.The suture was not returned.The needle was returned disengaged.Upon microscopic inspection, instrument marks were noted near the swaged end of the needle.Suture material was observed to be present in the swage, indicating the suture broke off at the swage.It was reported that the needle had detached.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The needle should always be held in the middle where the diameter was greatest.Grasping the needle near the swaged end or the tip could cause bends, breaks, or damage the connection between the needle and suture.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|