MEDOS INTERNATIONAL SÃ RL CH VIPER PRIME CFX FEN X-TAB POLYAXIAL SCREW 5.5 7 X 35MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
|
Back to Search Results |
|
| Catalog Number |
186770435S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Pain (1994); Numbness (2415)
|
| Date of Event |
05/06/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had lumber decompression and spinal fixation surgery at l4-s1 in (b)(6) 2023.The patient came back to the surgery clinic for follow-up after suffering from severe lower limbs pain and numbness.The patient had mri based on surgeon request to diagnose and investigate for the reason, and the surgeon found the bilateral screws at s1 had been broken, so immediately the surgeon admitted the patient for revision surgery to remove and replaced the broken screws with new ones.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: a2 (age), d6a, d6b and d10.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9 h3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: corrected: h4.H3, h6 photo investigation: the photo was returned to j&j medtech orthopaedics for evaluation.The j&j medtech orthopaedics team conducted a visual inspection of the returned photo.Visual analysis of the photo revealed that the viper prime cfxfn xtb 7x35mm s is seen broken, all pieces are seen in attached evidence.While no root cause can be established with the available information, factors such as the patient's age, underlying disease, bone density, or a possible early weight-bearing, could have led to implant failure.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for viper prime cfxfn xtb 7x35mm s.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Physical product investigation: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found it to be broken from the screw close to the screw head.The broken fragment was returned for evaluation.While no root cause can be established with the available information, factors such as the patient's age, underlying disease, bone density, or a possible early weight-bearing, could have led to implant failure.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test wass not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the viper prime cfxfn xtb 7x35mm s would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device product code: 186770435s lot number: tbaigi it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 27 jan 2022 manufacturing site: jabil le locle expiry date: 31 dec 2026.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|