Model Number 24690
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and severely calcified vessel below the knee.A 3.0mm x 40mm x 135cm sterling mr balloon catheter was advanced for dilation.During the procedure, the balloon ruptured upon first inflation at around 14 atmospheres for about 30 seconds.The device was removed without any problem, and the procedure was completed with a different device.There was no patient complications noted as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key22094734
Report Number2124215-2025-34612
Device Sequence Number7866450
Product Code DQY
UDI-Device Identifier08714729756880
UDI-Public(01)08714729756880(17)270422(10)33871294
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141150
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24690
Device Catalogue Number24690
Device Lot Number0033871294
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2025
Initial Report FDA Received Date05/28/2025
Date Device Manufactured04/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE - JUPITER FC; INFLATION DEVICE - GM30; INTRODUCER SHEATH - PARENT 4.5FR
Patient SexUnknown
-
-