• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RAYNER INTRAOCULAR LENSES LIMITED RAYONE ASPHERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RAYNER INTRAOCULAR LENSES LIMITED RAYONE ASPHERIC Back to Search Results
Model Number RAO600C
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 15th may 2025, rayner received notification from a us healthcare facility of an event that occurred during use of a rayone aspheric rao600c.The event description provided states that the lens was torn during insertion necessitating lens explantation and exchange.
 
Additional Manufacturer Narrative
The reference (b)(4) has been allocated to this case by rayner.The event description provided states that the lens tore during injection into the eye.The iol was explanted and exchanged during the original surgery session without further incident and without harm or injury to the patient."iol replacement or extraction" is listed in the "adverse events" section of the rayone ifu.The device has been discarded by the healthcare facility.The rayone preloaded iol injection system use risk analysis identifies the following as possible causes of "trapped/ torn lens haptic/ optic during insertion"; inadequate amount of viscoelastic; inadequate quality of viscoelastic; haptic trapped by plunger override due to fast motion; user opens closed flaps and closes again before use; plunger advanced too quickly, insertion of viscoelastic through nozzle leading to inadequate amount of viscoelastic; user removed injector from tray prior to inserting viscoelastic, causing lens to be improperly placed in cartridge; user removed injector from tray prior to closing cartridge, resulting in cartridge not being clipped properly; optic edge trapped/ damaged on closure of cartridge, sharp edge instruments and dehydration of the lens prior to implantation.Our review of production records for the rayone aspheric rao600c batch 054242517 showed that all manufacturing and quality checks were conducted with successful results.All devices released for distribution from this batch were within tolerance, met specification criteria and were without defects.A review of vigilance data confirms this is an isolated event.No other incidents, of any type, have been received against the rayone aspheric rao600c batch 054242517.There is insufficient evidence and information available to determine the cause of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAYONE ASPHERIC
Common Device Name
RAYONE ASPHERIC
Manufacturer (Section D)
RAYNER INTRAOCULAR LENSES LIMITED
the ridley innovation centre
10 dominion way
worthing, west sussex BN14 8AQ
UK  BN14 8AQ
Manufacturer (Section G)
RAYNER INTRAOCULAR LENSES LIMITED
the ridley innovation centre
10 dominion way
worthing, west sussex BN14 8AQ
UK   BN14 8AQ
Manufacturer Contact
jodie neal
the ridley innovation centre
10 dominion way
worthing, west sussex BN14 -8AQ
UK   BN14 8AQ
MDR Report Key22094776
Report Number3012304651-2025-00143
Device Sequence Number8022411
Product Code HQL
UDI-Device Identifier05029867692161
UDI-Public05029867692161
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P060011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date (Section B) 05/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/19/2026
Device Model NumberRAO600C
Device Catalogue NumberRAO600C
Device Lot Number054242517
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2025
Initial Report FDA Received Date05/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-