It was reported, during preparation of an ncompass nitinol tipless stone extractor, the basket was unable to open and close during an unspecified procedure.It was found that the yellow sheath and handle parts were separated.The procedure was completed by using another same-like device.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: h6 - annex a.Investigation ¿ evaluation.It was reported, during preparation of an ncompass nitinol tipless stone extractor, the basket was unable to open and close during an unspecified procedure.It was found that the yellow sheath and handle parts were separated.The procedure was completed by using another same-like device.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturer¿s instructions (mi), and quality control (qc) procedures were conducted during the investigation.The complaint device was not returned; therefore, no functional testing or visual inspections could be performed.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no related non-conformances reported for this failure mode.A complaint history search identified other related events for this failure mode.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in the field.Cook also reviewed product labeling; the ifu supplied with the device states the following in consideration of the reported failure mode: "precaution: the device is conductive.Avoid contact with any electrified instrument.Precaution: enclose the device in the sheath before removing from the tray/holder.Precaution: do not use excessive force to manipulate this device.Damage to the device may occur." based upon the available information, no product returned, and results of the investigation, cook has concluded that the cause of the event could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|