| Model Number |
2112 |
| Medical Device Problem Codes |
Degraded (1153); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3.Reason device not evaluated by mfg: other; device was not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the device had a constant cassette error, unable to clear.There was no reported patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
D3 - mfg establishment name- corrected.The suspected device was returned for evaluation.The event history log (ehl) was reviewed.The device was visually inspected and the downstream occlusion sensor was damaged.A functional test was performed and the reported issue was confirmed.The cause of the reported issue was due to the downstream occlusion sensor.As a result, the downstream occlusion sensor was replaced.The service history review had no indication that the complaint was related to a service of the device within the review period.The device passed all functional testing after repair and was returned to the customer.
|
| |
|