ENDOLOGIX SANTA ROSA OVATION IX; ILIAC LIMB Back to Search Results
Model Number TV-IL1422100-J
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 05/08/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was treated for an abdominal aortic aneurysm (aaa) with the implantation of the alto abdominal stent graft system on (b)(6) 2023.During a routine follow-up on (b)(6) 2025, the patient was reported to have a possible type ib endoleak.The previous year, the patient had been identified with a type ii endoleak (anatomy-related).The patient is reportedly in stable condition and is currently being monitored while an intervention is being discussed.
 
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows that the type 2 endoleak and possible type 1b endoleak are unconfirmed.This is not consistent with the reported adverse event/incident.Device, user, procedure or anatomy relatedness of complaint could not be determined.Procedure related harms could not be determined.The final patient status was reported as being monitored.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g3: awareness date has been updated h6: investigation finding codes: remove code 3233 h6: investigation conclusion codes: remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION IX
Common Device Name
ILIAC LIMB
Manufacturer (Section D)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
santa rosa CA 95403
Manufacturer (Section G)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
santa rosa CA 95403
Manufacturer Contact
gary kirchgater
3910 brickway blvd
santa rosa, CA 95403
8009832284
MDR Report Key22096675
Report Number3008011247-2025-00044
Device Sequence Number7866835
Product Code MIH
UDI-Device Identifier00850007370329
UDI-Public(01)00850007370329(17)250119(10)FS011922-16
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/19/2025
Device Model NumberTV-IL1422100-J
Device Lot NumberFS011922-16
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/08/2025
Supplement Date Received by Manufacturer07/16/2025
Initial Report FDA Received Date05/28/2025
Supplement Report FDA Received Date07/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTO MAIN BODY LOT# FS100722-25.; CUSTOMSEAL KIT FILL POLYMER LOT# FF081522-01.; OVATION IX ILIAC LIMB LOT# FS030422-03.
Outcome Attributed to Adverse Event Other;
Patient Age84 YR
Patient SexMale
-
-