The following information was reported to gore: on (b)(6) 2025, a gore® viabahn® vbx balloon expandable endoprosthesis (vbx device) was intended for use in the iliac artery during treatment of an aneurysm.After the device was advance into position the patient moved.The physician wanted to remove the vbx device before he ran another angiogram with contrast.During the removal of the vbx device through an 8fr x 10cm terumo introducer sheath, the vbx device dislodged from the delivery catheter.The patient was moved to the or and a cut down was performed in order to remove the undeployed vbx device.The physician then completed the procedure with a new vbx device.
|
|
Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.Manufacturing records were reviewed, and the device met all pre-release specifications.The device was discarded and was, therefore, not available for analysis.No clinical images enabling direct assessment of product performance were returned for evaluation.Cause of the reported event cannot be established based on the information reported to gore.Further information regarding this event was requested by gore, but no further information has been reported, therefore this investigation is considered complete.The gore® viabahn® vbx balloon expandable endoprosthesis instructions for use (ifu), for the appropriate region and time-period, was reviewed with respect to the complaint detail.The ifu includes the following warning: ¿do not withdraw the gore® viabahn® vbx balloon expandable endoprosthesis back into the introducer sheath once the endoprosthesis is fully introduced.Withdrawing the gore® viabahn® vbx balloon expandable endoprosthesis back into the introducer sheath can cause dislocation and/or dislodgement and damage to the endoprosthesis, premature deployment, deployment failure, and/or catheter separation which may result in vessel damage and/or ischemia, potentially requiring surgical intervention.If removal prior to deployment is necessary, withdraw the gore® viabahn® vbx balloon expandable endoprosthesis to a position close to but not into the introducer sheath.Both the gore® viabahn® vbx balloon expandable endoprosthesis and introducer sheath can then be removed in tandem.After removal, do not reuse the gore® viabahn® vbx balloon expandable endoprosthesis or introducer sheath.¿ w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|