Model Number DRIVER SHAFT, T4, NON-RETAINING, MINI AO
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 05/06/2025, a sales representative reported via email that the tip of the ar-18700-29 driver shaft broke off in the head of the ar-18714v-06 val screw as the surgeon was advancing it into the ar-18714p-21 hook plate.The broken pieces were retrieved from the patient.The ar-18700-29 driver shaft was being used with the jeweler's knob handle ar-18700h.The surgeon wanted to replace the screw, but the only way to remove it was to cut the tines off the ar-18714p-21 hook plate, remove each tine from the patient, and rotate the plate counterclockwise to back out the locking screw.The case was completed, and no further details were provided.This was discovered during an orif middle phalanx fracture procedure on (b)(6) 2025, with no reported patient harm.Additional information has been requested on 05/27/2025.
 
Event or Problem Description
Additional information was received on (b)(6) 2025: the broken shaft tip of the ar-18700-29 driver shaft was found inside the head of the ar-18714v-06 val screw that was removed.It remains uncertain whether the screw head became stripped due to the issue, as the driver tip was lodged inside.The tines of the ar-18714p-21 hook plate were successfully cut without complications, and no unplanned incisions resulted from the incident.No x-rays or photos were taken to document the event.There was no reported case delay, and no additional anesthesia was required.The patient's health status is stable, with no adverse effects observed during or after surgery.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.One unpackaged ar-18700-29 driver shaft fragment was received for investigation within an ar-18714v-06 val screw.Visual evaluation of the returned fragment noted it was stuck within the head of an ar-18714v-06 val screw.Functional testing was not able to be performed due to the damage to the device.The most likely cause for the reported failure can be attributed to overstressing a device that is damaged naturally and inevitably as a result of normal wear or aging from repeated usage/reprocessing.Manufactured date: 08-oct-2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVER SHAFT, T4, NON-RETAINING, MINI AO
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22097928
Report Number1220246-2025-02262
Device Sequence Number7343193
Product Code LXH
UDI-Device Identifier00888867303553
UDI-Public(01)00888867303553(11)191008(10)1391940
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDRIVER SHAFT, T4, NON-RETAINING, MINI AO
Device Catalogue NumberAR-18700-29
Device Lot Number1391940
Was Device Available for Evaluation? Device Returned to Manufacturer
Distributor Facility Aware Date05/06/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2025
Supplement Date Received by ManufacturerNot provided
09/30/2025
Initial Report FDA Received Date05/28/2025
Supplement Report FDA Received Date06/23/2025
10/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1