• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE MECHANICAL LITHOTRIPTOR V

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE MECHANICAL LITHOTRIPTOR V Back to Search Results
Model Number BML-V437QR-30
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the silicone rail into which the guide wire is inserted into on the lithotriptor was torn.The reported issue occurred during set up and there were no reports of patient harm.
 
Additional Manufacturer Narrative
The device was not returned to olympus for inspection; therefore, the customer's reported issue could not be confirmed.Based on the results of past investigations, it is likely the reported issue occurred due to the following mechanism: 1) an attempt was made to insert the device into the endoscope while using the guide wire.2) when the angle between the distal end and the guide wire was large as shown in the following figure, the device was inserted into the endoscope.3) a force exceeding the resisting force was applied to the guide wire tip.This caused the guide wire tip to tear.However, as the subject device was not returned, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE MECHANICAL LITHOTRIPTOR V
Common Device Name
SINGLE USE MECHANICAL LITHOTRIPTOR
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22098646
Report Number9614641-2025-00821
Device Sequence Number12850529
Product Code LQC
UDI-Device Identifier04953170462610
UDI-Public04953170462610
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
CLASS2-EXMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBML-V437QR-30
Device Lot Number48K
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/08/2025
Initial Report FDA Received Date05/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
G 260 2545S (SERIAL (B)(6)).
Patient SexUnknown
-
-